![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 137](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758137.webp)
CONFigure-Subsystem FSE
1065.6016.12
6.80
E-16
:CONFigure[:MS]:LIMit:PERCentile<numeric_value>
This command defines the limit for the 95% percentile.
Example:
":CONF:LIM:PERC 30"
Features:
*RST value
depending on the standard
SCPI:
device-specific
Mode:
BTS
:CONFigure[:MS]:LIMit:FREQuency <numeric_value>
This command determines the frequency error limits in ppm for the phase/frequency measurement.
Example:
":CONF:LIM:FREQ 36"
Feature:
*RST value:
depending on standard
SCPI:
device-specific
Mode:
MS
:CONFigure[:MS]:LIMit:STANdard ON | OFF
This command switches between user-defined (OFF) and standard-defined (ON) limit values.
Example:
":CONF:LIM:STAN ON"
Feature:
*RST value:
ON
SCPI:
device-specific
Mode:
MS
:CONFigure[:MS]:POWer:CLASs <numeric_value> | EG1 | EG2 | EG3
This command defines the power class of the mobile.
Parameter:
<numeric_value> ::=
1 to 5 (P-GSM phase I)
2 to 5 (P-GSM phase II, R-GSM850)
2 to 5 (E-GSM; R-GSM)
1 to 2 (DCS1800 phase I)
1 to 3 (DCS1800 phase II/II+)
1 to 3 (PCS1900)
Example:
":CONF:POW:CLAS 4"
Features:
*RST value:
2
(P-GSM phase I/II, E-GSM, R-GSM, R-GSM850)
1
(DCS1800, PCS1900)
SCPI:
device-specific
Mode:
MS
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......