SENSe Subsystem FSE
1065.6016.12
6.198
E-16
:[SENSe<1|2>:]MIXer:PORTs 2 | 3
This command activates the 2- or 3-port mixer. In the BAND LOCK ON mode, the command refers
to the active band selected with
SENSe:MIXer:HARMonic:BAND
.
Example:
":MIX:PORT 3"
Features:
*RST value:
2
SCPI:
device-specific
Mode:
A
This command is available only if the external mixer (option FSE-B21) is switched on.
:[SENSe<1|2>:]MIXer:SIGNal ON | OFF | AUTO
This command activates the Signal ID or Auto ID mode.
Example:
":MIX:SIGN ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A
This command is available only if the external mixer (option FSE-B21) is switched on.
:[SENSe<1|2>:]MIXer:THReshold 0.1 to 100 dB
This command sets the level threshold for auto ID.
Example:
":MIX:THR 20"
Features:
*RST value:
10
SCPI:
device-specific
Mode:
A
The command is available only if the external mixer (option FSE-B21) is switched on.
:[SENSe<1|2>:]MIXer:HARMonic 1 to 62
With BAND LOCK OFF, this command sets the nth harmonic. The command may be a query with
BAND LOCK ON.
Example:
":MIX:HARM 5"
Features:
*RST value:
2
SCPI:
conforming
Mode:
A
The command is available only if the external mixer (option FSE-B21) is switched on.
Summary of Contents for 1065.6000.20
Page 3: ......
Page 5: ......
Page 17: ......
Page 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Page 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Page 391: ......
Page 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Page 419: ......
Page 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Page 429: ......