![Rohde & Schwarz 1065.6000.20 Operating Manual Download Page 249](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758249.webp)
SENSe Subsystem FSE
1065.6016.12
6.192
E-16
:[SENSe<1|2>:]FILTer:DEMPhasis:TCONstant 50US | 75US | 750US
This command sets the time constant of the de-emphasis for analog demodulation.
Example:
":FILT:DEMP:TCON 75US"
Features:
*RST value:
50us
SCPI:
conforming
Mode:
VA-A
:[SENSe<1|2>:]FILTer:DEMPhasis:LINK DISPlay | AUDio
For analog demodulation, this command selects whether the de-emphasis set is to be active in the
audio branch only or in addition for the display of measured values.
Example:
":FILT:DEMP:LINK DISP"
Features:
*RST value:
AUDio
SCPI:
device-specific
Mode:
VA-A
AUDio
De-emphasis effective in the audio branch only
DISPlay
De-emphasis effective in the audio branch and in the display of measured values