404
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
25
CAN/LIN Triggering and Serial Decode
The DLC field will always be displayed in decimal, and will indicate the number
of bytes in the frame. So for example, for an FD frame that has the DLC code
0xF, which represents a packet with 64 bytes, "DLC=64" will be displayed on
the decode line, and "64" will be displayed in the DLC column of the Lister.
• Overload frame — blue with the text "OVRLD". An overload condition may occur
before an end of frame condition. If so, the frame is closed and opened with
blue brackets at the start of the overload condition.
• Stuff Count — green hex digit when valid, red when error detected. The hex digit
shows the gray-coded with parity bit stuff count.
• CRC — blue hex digits when valid, red when error detected.
• Red angled waveforms — Unknown or error condition.
• Flagged error frames — red with "ERR FRAME", "STUFF ERR", "FORM ERR",
"ACK ERR", "GLITCH ERR", or "?" (unknown).
• Pink vertical bars — Expand horizontal scale (and run again) to see decode.
• Red Dot — More information is available. Decoded text is truncated to fit.
Expand the horizontal scale to view the information.
CAN Totalizer
The CAN totalizer provides a direct measure of bus quality and efficiency. The CAN
totalizer measures total CAN frames, Specification error counter, flagged error
frames, and bus utilization.
The totalizer is always running (counting frames and calculating percentages) and
is displayed whenever CAN decode is displayed. The totalizer counts even when
the oscilloscope is stopped (not acquiring data). Pressing the
[Run/Stop]
key does
not affect the totalizer. When an overflow condition occurs, the counter displays
OVERFLOW
. The counters can be reset to zero by pressing the
Reset CAN Counters
softkey.
Types of Frames
• Active error frames are CAN frames in which a CAN node recognizes an error
condition during a data or remote frame and issues an active error flag.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...