SENT Triggering and Serial Decode
32
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
503
•
Successive Sync Pulses Error
— triggers on a sync pulse whose width varies
from the previous sync pulse's width by greater than 1/64 (1.5625%, as
defined in the SENT specification).
4
If you choose the
Fast Channel SC & Data
trigger condition:
a
Press the
Base
softkey to choose between Hex or Binary data value entry.
Use the Binary method if you need to enter "don't care" bits (X) within a
nibble. If all bits in a nibble are "don't care", the hex nibble is displayed as
"don't care" (X). When all bits in a nibble are 1s or 0s, the hex value is shown.
Hex nibbles that contain a mix of 0/1 bits and "don't care" bits are displayed
as "$".
b
Press the
SC & Data
softkey and use the keypad dialog to enter the data
value.
The S&C nibble is the left most nibble entered in the string of numbers,
followed by the data nibbles.
5
If you choose the
Slow Channel Message ID
or
Slow Channel Message ID & Data
trigger condition:
a
Press the
Configuration
softkey to select the desired packet ID for the selected
packet type.
When an Enhanced Serial message format is specified (see
on page 497), press this softkey to select which enhanced format
configuration is being used:
•
16-Bit Data, 4-Bit ID
•
12-Bit Data, 8-Bit ID
b
Press the
Slow Message ID
softkey and use the Entry knob (or press the
softkey again and use a keypad dialog) to specify the Slow Message ID.
c
If you choose the
Slow Channel Message ID & Data
trigger condition, press the
Slow Data
softkey and use the Entry knob (or press the softkey again and use
a keypad dialog) to specify the Slow Message data.
6
If you choose the
Tolerance Violation
trigger condition, press the
Tolerance
softkey and use the Entry knob (or press the softkey again and use a keypad
dialog) to specify the tolerance variation that is considered a violation.
The percentage entered must be less than the percent tolerance specified in
the decode bus configuration settings.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...