When triggering on a positive pulse, the trigger will occur on the high to low
transition of the pulse if the qualifying condition is true. When triggering on a
negative pulse, the trigger will occur on the low to high transition of the pulse if
the qualifying condition is true.
6
Press the qualifier softkey (
< > ><
) to select the time qualifier.
The Qualifier softkey can set the oscilloscope to trigger on a pulse width that is:
•
Less than a time value (
<
).
For example, for a positive pulse, if you set t<10 ns:
•
Greater than a time value (
>
).
For example, for a positive pulse, if you set t>10 ns:
•
Within a range of time values (
><
).
For example, for a positive pulse, if you set t>10 ns and t<15 ns:
7
Select the qualifier time set softkey (
<
or
>
), then rotate the Entry knob to set
the pulse width qualifier time.
The qualifiers can be set as follows:
•
2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz and lower
bandwidth models).
•
10 ns to 10 s for >< qualifier, with minimum difference of 5 ns between upper
and lower settings.
Pulse width
trigger < qualifier
time set softkey
• When the less than (
<
) qualifier is selected, the Entry knob sets the oscilloscope
to trigger on a pulse width less than the time value displayed on the softkey.
10 ns
10 ns
Trigger
10 ns
10 ns
Trigger
10 ns
15 ns
Trigger
12 ns
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...