Digital Voltmeter and Counter
2
Press
Features
; then, select
Counter
.
3
Press
Features
again to enable the counter.
4
Press the
Source
softkey and turn the Entry knob to select the analog channel or
Trigger Qualified Event
signal to make counter measurements on.
With the
Trigger Qualified Event
source (available when the trigger mode is not
Edge), you can see how often trigger events are detected. This can be more
often than when triggers actually occur, due to the oscilloscope's acquisition
time or update rate capabilities. The TRIG OUT signal shows when triggers
actually occur. Remember that the oscilloscope's trigger circuitry does not
re-arm until the holdoff time occurs and that the minimum holdoff time is 40 ns;
therefore, the maximum trigger qualified event frequency that can be counted
is 25 MHz.
The selected channel does not have to be on (displaying a waveform) in order
for counter measurements to be made.
5
Press
Auto Setup Threshold
softkey to have the oscilloscope automatically
determine and set the threshold voltage (trigger) level for the selected analog
channel source.
6
Press the
Measure
softkey and turn the Entry knob to select what the counter
measures:
•
Frequency
— the cycles per second (Hz, kHz, or MHz) of the signal.
•
Period
— the time periods of the signal's cycles.
•
Totalize
— the count of edge events on the signal.
Frequency and
Period Counter
For frequency and period measurements, press the
# of Digits
softkey to specify the
resolution of the counter. You can choose from 3 to 8 digit resolutions.
Higher resolutions require longer gate times, which cause the measurement times
to be longer as well.
Totalize Counter
For (edge event) totalize measurements, press the
Clear Count
softkey to zero the
edge event counter.
Press the
Totalize
softkey to open the Counter Totalize Menu where you can:
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...