Digital Channels
7
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
145
The impedance plots for the two models are shown in these figures. By comparing
the two plots, you can see that both the series tip resistor and the cable's
characteristic impedance extend the input impedance significantly. The stray tip
capacitance, which is generally small (1 pF), sets the final break point on the
impedance chart.
The logic probes are represented by the high-frequency circuit model shown
above. They are designed to provide as much series tip resistance as possible.
Stray tip capacitance to ground is minimized by the proper mechanical design of
the probe tip assembly. This provides the maximum input impedance at high
frequencies.
Probe Grounding
A probe ground is the low-impedance path for current to return to the source from
the probe. Increased length in this path will, at high frequencies, create large
common mode voltages at the probe input. The voltage generated behaves as if
this path were an inductor according to the equation:
Figure 24
Impedance versus Frequency for Both Probe Circuit Models
100 k
10 k
1 k
100
10
1
10 kHz
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
High
Frequency
Model
Typical
Model
Frequency
Impedance
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...