112
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
5
Math Waveforms
4
To make cursor measurements, press the
[Cursors]
key and set the
Source
softkey to
Math N
.
Use the X1 and X2 cursors to measure frequency values and difference between
two frequency values (
Δ
X). Use the Y1 and Y2 cursors to measure amplitude in
dB and difference in amplitude (
Δ
Y).
5
To make other measurements, press the
[Meas]
key and set the
Source
softkey to
Math N
.
You can make peak-to-peak, maximum, minimum, and average dB
measurements on the FFT waveform. You can also find the frequency value at
the first occurrence of the waveform maximum by using the X at Max Y
measurement.
The following FFT (Magnitude) spectrum was obtained by connecting a 2.5 V,
100 kHz square wave to channel 1. Set the horizontal scale to 50 µs/div, vertical
sensitivity to 1 V/div, Units/div to 20 dBV, Offset to -40.0 dBV, Center frequency to
500 kHz, frequency Span to 1 MHz, and window to Hanning.
NOTE
Scale and offset considerations
If you do not manually change the FFT scale or offset settings, when you turn the horizontal
scale knob, the span and center frequency settings will automatically change to allow
optimum viewing of the full spectrum.
If you do manually set scale or offset, turning the horizontal scale knob will not change the
span or center frequency settings, allowing you see better detail around a specific frequency.
Pressing the FFT
Auto Setup
softkey will automatically rescale the waveform and span and
center will again automatically track the horizontal scale setting.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...