146
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
7
Digital Channels
Increasing the ground inductance (L), increasing the current (di) or decreasing the
transition time (dt), will all result in increasing the voltage (V). When this voltage
exceeds the threshold voltage defined in the oscilloscope, a false data
measurement will occur.
Sharing one probe ground with many probes forces all the current that flows into
each probe to return through the same common ground inductance of the probe
whose ground return is used. The result is increased current (di) in the above
equation, and, depending on the transition time (dt), the common mode voltage
may increase to a level that causes false data generation.
Figure 25
Common Mode Input Voltage Model
V
=
L
di
dt
Z
in
Z
in
i
n
i
2
+i
2
i
1
i
1
+i
n
+i
n
Z
in
Probe 1
Probe
Ground
Probe 2
Probe N
L (GND)
Vn (Common Mode)
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...