Utility Settings
22
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
351
•
Trigger Source
— The raw trigger signal from the oscilloscope's trigger circuit
is output to Trig Out. It produces a rising edge whenever the input source
would cause a trigger, even though that might occur multiple times within
the time of a single acquisition. The trigger source can be a front panel
analog input channel or an external trigger input. The output level is 0-5 V
into an open circuit, and 0-2.5 V into 50
Ω
. This option is not available in all
trigger modes.
The TRIG OUT connector also provides the User Calibration signal. See
Enabling Remote Command Logging
When remote command logging is enabled, remote commands sent to the
instrument (and results returned by the instrument) can be logged to the screen,
to a text file on a USB storage device, or to both the screen and a text file.
To enable remote command logging:
1
Press
[Utility] > Options > Remote Log
to open the Remote Log Menu:
2
Press
Enable
to enable or disable the remote command logging feature.
When remote logging is enabled, additional debug information can be included
in the returned error string. If the error is detected by the SCPI command
parser, such as a header error or other syntax error, the extra debug
information is generated and included. But if the error is detected by the
oscilloscope system, such as when an out-of-range value is sent, then no extra
debug information is included.
3
Press
Destination
to select whether remote commands are logged to a text file
(on a connected USB storage device), logged to the screen, or both.
4
Press
Write Mode
to specify whether logged commands will be created in a new
list or appended to existing logged commands.
Your selection takes effect when remote command logging is enabled.
This option applies to both screen and file logging.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...