Measurements
15
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
259
Counter
The InfiniiVision X-Series oscilloscopes have an integrated hardware frequency
counter which counts the number of cycles that occur within a period of time
(known as the gate time) to measure the frequency of a signal.
The gate time is the horizontal range of the oscilloscope but is limited to >= 0.1 s
and <= 10 s. Unlike other measurements, the Zoom horizontal timebase window
does not gate the Counter measurement.
The Counter measurement can measure frequencies up to the bandwidth of the
oscilloscope. The minimum frequency supported is 2.0 / gateTime.
The hardware counter uses the trigger comparator output. Therefore, the counted
channel's trigger level (or threshold for digital channels) must be set correctly.
Analog and digital channels can be selected as the source.
Only one Counter measurement can be displayed at a time.
Figure 49
Isolating an event for Frequency measurement
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...