152
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
8
Serial Decode
Before you can select a row or navigate through the Lister data, oscilloscope
acquisitions must be stopped.
5
Press the
[Single]
key (in the Run Control group on the front panel) to stop the
acquisition.
Pressing
[Single]
instead of
[Stop]
fills the maximum memory depth.
When zoomed out and viewing a large number of packets, the Lister may not
be able to display information for all packets. However, when you press the
[Single]
key the Lister will contain all on-screen serial decode information.
6
Press the
Scroll Lister
softkey and use the Entry knob to scroll through the data.
Time tags in the Time column indicate the event time relative to the trigger
point by default, and can optionally be configured to be relative to the previous
row, as described in step 9 that follows. The time tags of events that are shown
in the waveform display area are displayed with a dark background.
7
Press the
Zoom to Selection
softkey (or push the Entry knob) to center the
waveform display at the time associated with the selected Lister row and
automatically set the horizontal scale setting.
8
Press the
Undo Zoom
softkey to return to the horizontal scale and delay settings
before the last
Zoom to Selection
.
9
Press the
Options
softkey to open the Lister Options Menu. In this menu, you
can:
•
Enable or disable the
Track Time
option. When enabled, as you select
different Lister rows (using the Entry knob while acquisitions are stopped),
the horizontal delay changes to the Time of the selected row. Also, changing
the horizontal delay will scroll the Lister.
•
Press the
Scroll Lister
softkey and use the Entry knob to scroll though data
rows in the Lister display.
•
Press the
Time Ref
softkey and use the Entry knob to select whether the Time
column in the Lister display shows times relative to the trigger or relative to
the previous packet row.
Searching Lister Data
When serial decode is enabled, you can use the
[Search]
key to find and place
marks on rows in the Lister.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...