Acquisition Control
13
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
229
When Realtime sampling is on, the bandwidth of the oscilloscope is limited
because the bandwidth of the reconstruction filter is set to fs/4. For example, a
MSOX410xA oscilloscope with channels 1 and 2 on has a bandwidth of 625 MHz
when Realtime sampling on and 1 GHz when Realtime sampling is off.
The sample rate is displayed in the sidebar Summary dialog.
Acquiring to Segmented Memory
When capturing multiple infrequent trigger events it is advantageous to divide the
oscilloscope's memory into segments. This lets you capture signal activity without
capturing long periods of signal inactivity.
Each segment is complete with all analog channel, digital channel (on MSO
models), and serial decode data.
When using segmented memory, use the Analyze Segments feature (see
“Measurements, Statistics, and Infinite Persistence with Segmented Memory"
page 231) to show infinite persistence across all acquired segments. See also
To acquire to
segmented
memory
1
Set up a trigger condition. (See
, “Triggers,” starting on page 169
for details.)
2
Press the
[Acquire]
key in the Waveform section of the front panel.
3
Press the
Segmented
softkey.
4
In the Segmented Memory Menu, press the
Segmented
softkey to enable
segmented memory acquisitions.
5
Press the
# of Segs
softkey and turn the Entry knob to select the number of
segments into which you would like to divide the oscilloscope's memory.
Memory can be divided into as few as two segments and as many as 1000
segments, depending on the oscilloscope model.
6
Press the
[Run]
or
[Single]
key.
The oscilloscope runs and fills a memory segment for each trigger event. When
the oscilloscope is busy acquiring multiple segments, the progress is displayed on
screen. The oscilloscope continues to trigger until memory is filled, then the
oscilloscope stops.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...