Manchester/NRZ Triggering and Serial Decode
30
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
473
•
Errors
Aliased data is highlighted in pink. When this happens, decrease the horizontal
time/div setting and run again.
Setup for NRZ Signals
NRZ signal setup consists of first connecting the oscilloscope to a NRZ-encoded
serial bus signal, then specifying the signal, bus configuration, and other settings.
To set up the oscilloscope to capture NRZ signals:
1
Press
[Serial]
.
2
Press the
Serial
softkey and choose the serial decode (Serial 1 or Serial 2) to
use; then, press the softkey again to enable decode.
3
Press the
Mode
softkey; then, select the
NRZ
decode mode.
4
Press the
Signals
softkey to open the NRZ Signals Menu.
In the NRZ Signals Menu:
a
Press the
Source
softkey to select the channel connected to the NRZ signal
line.
The label for the NRZ source channel is automatically set.
b
Press the
Threshold
softkey; then, enter the trigger threshold voltage level.
The threshold voltage level is used in decoding and will become the trigger
level when the trigger type is set to the NRZ serial decode.
c
Press the
Baud
softkey; then, enter the baud rate of the NRZ signal from your
device under test.
The baud rate can be set from 5 kb/s to 5 Mb/s in 100 b/s increments.
You must set the baud rate to match your device under test.
The default baud rate is 125 kb/s.
d
Press the
Back/Up key to return to the Serial Decode Menu.
Back
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...