Measurements
15
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
255
DC RMS
DC RMS is the root-mean-square value of the waveform over one or more full
periods.
Where x
i
= value at
i
th point being measured, n = number of points in
measurement interval.
The Full Screen measurement interval variation measures the value on all
displayed data points.
The N Cycles measurement interval variation measures the value on an integral
number of periods of the displayed signal. If less than three edges are present, the
measurement shows "No edges".
The X cursors show the interval of the waveform being measured.
AC RMS
AC RMS is the root-mean-square value of the waveform, with the DC component
removed. It is useful, for example, for measuring power supply noise.
The N Cycles measurement interval measures the value on an integral number of
periods of the displayed signal. If less than three edges are present, the
measurement shows "No edges".
The X cursors show the interval of the waveform being measured.
The Full Screen (Std Deviation) measurement interval variation is an RMS
measurement across the full screen with the DC component removed. It shows the
standard deviation of the displayed voltage values.
The standard deviation of a measurement is the amount that a measurement
varies from the mean value. The Mean value of a measurement is the statistical
average of the measurement.
RMS (dc) =
∑
i
=1
n
x
i
2
n
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...