SENT Triggering and Serial Decode
32
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
507
The decode will also indicate errors where a nibble's pulse is too wide or too
narrow. This will be indicated by a ">" or "<" in red, as well as the rest of the
packet's outline and closing brace being in red, as well as an orange idle line until
the next valid sync. At the time of the valid sync, there will be an orange opening
brace.
Interpreting SENT Lister Data
Each Fast or Slow Channel message is represented on its own row. The start time
of Slow Channel messages determines their order relative to the Fast Channel
messages. Therefore, a Slow Channel message appears before most of the Fast
Channel messages that it was built from. This is due to the "Time" column holding
the start time of a packet.
In addition to the standard Time column, the following columns are used to
support both the Fast Channel and the Slow Channel simultaneously, and these
columns are displayed for all Message Format modes except
Fast Nibbles (All)
:
• S&C — (Fast Channel Only) (binary).
• ID — (Slow Channel Only) (Hex or Decimal).
• Data — (Hex or Decimal):
•
Fast Channel:
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...