Triggers
11
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
185
3
Press the
Source
softkey, and turn the Entry knob to select an analog input
channel source.
4
Press the
Standard
softkey to select the signaling technology used by the input
signal.
5
Set the trigger level to capture transitions in the modulation envelope. Or at
approximately 50% of the modulation depth. NFC-A uses 100% modulation, so
the trigger level can be set in a wide range of values. NFC-B, NFC-F212, and
NFC-F424 use 10% modulation, so the trigger level needs to be set to
approximately 95% of the unmodulated carrier amplitude. Usually the actual
modulation depth is greater than 10%, so the trigger level can be set
accordingly.
The following image shows how the trigger level should be set. The Y2 marker
is set for the unmodulated carrier amplitude or 123.3 mV. The Y1 marker is set
for the maximum modulated amplitude or 97.8 mV. (Note that the actual
modulation depth is close to 20%.) The midpoint of the modulation would be
(123.3 mV + 97.8 mV) / 2.0 = 110.6 mV. The trigger level is set to 108 mV;
slightly below the mid modulation level to account for trigger hysteresis.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...