194
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
11
Triggers
The
Generic
selection lets you trigger on custom bi-level and tri-level sync video
“To set up Generic video triggers"
7
Press the
Auto Setup
softkey to automatically set up the oscilloscope for the
selected
Source
and
Standard
:
•
Source channel vertical scaling is set to 140 mV/div.
•
Source channel offset is set to 245 mV.
•
Source channel is turned on.
•
Trigger type is set to
Video
.
•
Video trigger mode is set to
All Lines
(but left unchanged if
Standard
is
Generic
).
•
Display
Grid
type is set to
IRE
(when
Standard
is
NTSC
) or
mV
(see
•
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM standards
or 4 µs/div for EDTV or HDTV standards (unchanged for
Generic
).
•
Horizontal delay is set so that trigger is at first horizontal division from the
left (unchanged for
Generic
).
You can also press
[Analyze]> Features
and then select
Video
to quickly access the
video triggering automatic set up and display options.
8
Press the
Mode
softkey to select the portion of the video signal that you would
like to trigger on.
The Video trigger modes available are:
•
Field1
and
Field2
— Trigger on the rising edge of the first serration pulse of
field 1 or field 2 (interlaced standards only).
•
All Fields
— Trigger on the rising edge of the first pulse in the vertical sync
interval.
•
All Lines
— Trigger on all horizontal sync pulses.
•
Line
— Trigger on the selected line number (EDTV and HDTV standards only).
•
Line: Field1
and
Line:Field2
— Trigger on the selected line # in field 1 or field 2
(interlaced standards only).
•
Line: Alternate
— Alternately trigger on the selected line # in field 1 and field 2
(NTSC, PAL, PAL-M, and SECAM only).
9
If you select a line # mode, press the
Line #
softkey, then rotate the Entry knob
to select the line number on which you want to trigger.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...