Acquisition Control
13
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
221
Memory Depth and Sample Rate
The number of points of oscilloscope memory is fixed, and there is a maximum
sample rate associated with oscilloscope's analog-to-digital converter; however,
the actual sample rate is determined by the time of the acquisition (which is set
according to the oscilloscope's horizontal time/div scale).
sample rate = number of samples / time of acquisition
For example, when storing 50 µs of data in 50,000 points of memory, the actual
sample rate is 1 GSa/s.
Likewise, when storing 50 ms of data in 50,000 points of memory, the actual
sample rate is 1 MSa/s.
The actual sample rate is displayed in the Summary box in the right-side
information area.
The oscilloscope achieves the actual sample rate by throwing away (decimating)
unneeded samples.
Selecting the Acquisition Mode
When selecting the oscilloscope acquisition mode, keep in mind that samples are
normally decimated at slower time/div settings.
At slower time/div settings, the effective sample rate drops (and the effective
sample period increases) because the acquisition time increases and the
oscilloscope's digitizer is sampling faster than is required to fill memory.
For example, suppose an oscilloscope's digitizer has a sample period of 1 ns
(maximum sample rate of 1 GSa/s) and a 1 M memory depth. At that rate, memory
is filled in 1 ms. If the acquisition time is 100 ms (10 ms/div), only 1 of every 100
samples is needed to fill memory.
To select the acquisition mode:
1
Press the
[Acquire]
key on the front panel.
2
In the Acquire Menu, press the
Acq Mode
softkey; then, turn the Entry knob to
select the acquisition mode.
The InfiniiVision oscilloscopes have the following acquisition modes:
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...