Triggers
11
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
201
oscilloscope will trigger on color field 1 alternating with color field 3 (see the
following figure). This setup can be used to measure the envelope of the reference
burst.
If a more detailed analysis is required, then only one color field should be selected
to be the trigger. You can do this by using the
Field Holdoff
softkey in the Video
Trigger Menu. Press the
Field Holdoff
softkey and use the Entry knob to adjust the
holdoff in half-field increments until the oscilloscope triggers on only one phase of
the color burst.
A quick way to synchronize to the other phase is to briefly disconnect the signal
and then reconnect it. Repeat until the correct phase is displayed.
When holdoff is adjusted using the
Field Holdoff
softkey and the Entry knob, the
corresponding holdoff time will be displayed in the Trigger Mode and Coupling
Menu.
Figure 32
Triggering on Color Field 1 Alternating with Color Field 3
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...