Trigger Mode/Coupling
12
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
209
Trigger Indicator
The trigger indicator at the top right of the display shows whether triggers are
occurring.
In the
Auto
trigger mode, the trigger indicator can show:
•
Auto?
— the trigger condition is not found (after the pre-trigger buffer has filled),
and forced triggers and acquisitions are occurring.
•
Auto
— the trigger condition is found (or the pre-trigger buffer is being filled).
In the
Normal
trigger mode, the trigger indicator can show:
•
Trig'd?
— the trigger condition is not found (after the pre-trigger buffer has
filled), and no acquisitions are occurring.
•
Trig'd
— trigger condition is found (or pre-trigger buffer is being filled).
When the oscilloscope is not running, the trigger indicator area shows
Stop
.
When to Use Auto
Trigger Mode
The
Auto
trigger mode is appropriate when:
• Checking DC signals or signals with unknown levels or activity.
• When trigger conditions occur often enough that forced triggers are
unnecessary.
When to Use
Normal Trigger
Mode
The
Normal
trigger mode is appropriate when:
• You only want to acquire specific events specified by the trigger settings.
• Triggering on an infrequent signal from a serial bus (for example, I2C, SPI, CAN,
LIN, etc.) or another signal that arrives in bursts. The
Normal
trigger mode lets
you stabilize the display by preventing the oscilloscope from auto-triggering.
• Making single-shot acquisitions with the
[Single]
key.
Often with single-shot acquisitions, you must initiate some action in the device
under test, and you don't want the oscilloscope to auto-trigger before that
happens. Before initiating the action in the circuit, wait for the trigger condition
indicator
Trig'd?
to display (this tells you the pre-trigger buffer is filled).
See Also
•
•
•
“To position the time reference (left, center, right, custom)"
To select the trigger coupling
1
Press the
[Mode/Coupling]
key.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...