468
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
30
Manchester/NRZ Triggering and Serial Decode
•
Falling — specifies that a falling edge is used to encode a bit value of logic
1 (and a rising edge encodes a bit value of logic 0).
c
Press the
Bit Order
softkey and select the bit order, either most significant bit
first (
MSB
) or least significant bit first (
LSB
).
The selected order is used when displaying data in the serial decode
waveform and in the Lister.
This softkey is available when the
Word
display format is selected in the
Manchester Bus Configuration Menu.
d
Press the
Idle Bits
softkey and enter the minimum idle time/inter-frame gap
time of the Manchester bus in terms of the bit width.
Framing is determined by idle time.
e
Use the
Decode Base
softkey to select between hexadecimal, unsigned
decimal, or ASCII display of the decoded data.
The decode base setting is used for the display of the data field in both the
decode line and in the Lister.
When the
Bit
display format is selected in the Manchester Bus Configuration
Menu, this softkey is unavailable and the decode base is forced to Binary.
f
Press the
Back/Up key to return to the Serial Decode Menu.
Manchester Triggering
To set up the oscilloscope to capture a Manchester signal, see
To set up a Manchester trigger:
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger Type
softkey; then, select the Manchester
decode (Serial 1 or Serial 2).
3
Press the
Trigger
softkey; then, select the trigger condition:
Back
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...