Digital Voltmeter and Counter
17
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
293
• Press the
Source
softkey and turn the Entry knob to change the analog channel
on which counter measurements are made.
• Press the
Event Slope
softkey to choose whether positive or negative edge
events are counted.
• Press the
Gate
softkey to enable or disable gating of the edge event count using
a positive or negative level on a second analog channel.
When gating is enabled:
a
Press the
Gate Source
softkey and turn the Entry knob to select the analog
channel that will supply the gating signal.
The selected channel does not have to be on (displaying a waveform).
b
Press the polarity softkey to choose whether positive or negative levels are
used to gate the edge event count.
The trigger level for the selected analog channel is used to determine the
polarity of the signal.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...