MIL-STD-1553/ARINC 429 Triggering and Serial Decode
31
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
483
MIL-STD-1553 Triggering
To set up the oscilloscope to capture a MIL-STD-1553 signal, see
To set up a MIL-STD-1553 trigger:
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select the serial slot (Serial 1 or Serial 2) on which the MIL-STD-1553 signal is
being decoded.
3
Press the
Trigger
softkey; then, turn the Entry knob to select the trigger
condition:
•
Data Word Start
— triggers on the start of a Data word (at the end of a valid
Data Sync pulse).
•
Data Word Stop
— triggers on the end of a Data word.
•
Command/Status Word Start
— triggers on the start of Command/Status word
(at the end of a valid C/S Sync pulse).
•
Command/Status Word Stop
— triggers on the end of a Command/Status word.
•
Remote Terminal Address
— triggers if the RTA of the Command/Status word
matches the specified value.
When this option is selected, the
RTA
softkey becomes available and lets you
select the hex Remote Terminal Address value to trigger on. If you select
0xXX (don't cares), the oscilloscope will trigger on any RTA.
•
Remote Terminal A 11 Bits
— triggers if the RTA and the remaining 11
bits match the specified criteria.
When this option is selected, these softkeys become available:
•
The
RTA
softkey lets you select the hex Remote Terminal Address value.
•
The
Bit Time
softkey lets you select the bit time position.
•
The
0 1 X
softkey lets you set the bit time position value as a 1, 0, or X
(don't care).
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...