How is mask testing done?
InfiniiVision oscilloscopes start mask testing by creating a database that is 200 x
640 for the waveform viewing area. Each location in the array is designated as
either a violation or a pass area. Each time a data point from a waveform occurs in
a violation area a failure is logged. If
Test All
was selected, every active analog
channel is tested against the mask database for each acquisition. Over 2 billion
failures can be logged per-channel. The number of acquisitions tested is also
logged and displayed as "# of Tests".
The mask file allows greater resolution than the 200 X 640 database. Some
quantization of data occurs to reduce the mask file data for display on-screen.
NOTE
Although a mask can have up to 8 regions, in any given vertical column, it is only possible to
define 4 regions. When there are 4 regions in a vertical column, one region must be tied to the
top (using the MAX y value) and one must be tied to the bottom (using the MIN y value).
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...