210
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
12
Trigger Mode/Coupling
2
In the Trigger Mode and Coupling Menu, press the
Coupling
softkey; then, turn
the Entry knob to select:
•
DC
coupling — allows DC and AC signals into the trigger path.
•
AC
coupling — places a 10 Hz high-pass filter in the trigger path removing
any DC offset voltage from the trigger waveform.
The high-pass filter in the External Trigger input path is 50 Hz for all models.
Use AC coupling to get a stable edge trigger when your waveform has a
large DC offset.
•
LF
(low frequency)
Reject
coupling — adds a high-pass filter with the 3-dB
point at 50 kHz in series with the trigger waveform.
Low frequency reject removes any unwanted low frequency components
from a trigger waveform, such as power line frequencies, etc., that can
interfere with proper triggering.
Use
LF Reject
coupling to get a stable edge trigger when your waveform has
low frequency noise.
•
Video
coupling — is normally grayed-out, but is automatically selected when
Video trigger is enabled in the Trigger Menu.
Note that Trigger Coupling is independent of Channel Coupling (see
Pass
Band
50 kHz
0 dB
3 dB down point
DC
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...