460
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
29
I2S Triggering and Serial Decode
If you select an "Armed" value that is equal to or greater than the "Trigger"
value, the "Trigger" value will be increased so that it is always greater than
the "Armed" value.
•
Decreasing value
— similar to the description above except the trigger occurs
on a decreasing data word value, and the "Armed" value is the value to
which the data must rise in order to re-arm the trigger.
6
Press the
Base
softkey and select a number base for entering data values:
•
Binary (2's complement)
.
When Binary is selected, the
Bits
softkey appears. This softkey opens the I2S
Bits Menu for entering data values.
When the trigger qualifier requires a pair of values (as with In Range, Out of
Range, Increasing value, or Decreasing value), the first softkey in the I2S Bits
Menu lets you select which value of the pair.
In the I2S Bits Menu, press the
Bit
softkey and rotate the Entry knob to select
each bit; then, use the
0 1 X
softkey to set each bit value to zero, one, or
don't care. You can use the
Set all Bits
softkey to set all bits to the value
chosen on the
0 1 X
softkey. Don't care values are only allowed with Equal or
Not Equal trigger qualifiers.
•
Signed decimal
.
When Decimal is selected, the softkey(s) to the right let you enter decimal
values with the Entry knob. These softkeys can be
Data
,
<
,
>
, or
Threshold
depending on the selected trigger qualifier.
"Trigger" value
Trigger
Trigger
No trigger
"Armed" value
NOTE
If the setup does not produce a stable trigger, the I2S signal may be slow enough that the
oscilloscope is AutoTriggering. Press the
[Mode/Coupling]
key, then press the
Mode
softkey
to set the trigger mode from
Auto
to
Normal
.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...