Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
20
Save/Email/Recall (Setups, Screens, Data)
When you want to save less than the maximum number of data points, a 1-of-N
decimation is performed to produce an output whose length is less than or equal
to the requested length. For example, if there are 100k points of data, and you
specify a length of 2k, 1 of every 50 data points is saved.
See Also
•
•
•
“Minimum and Maximum Values in CSV Files"
Length Control
The
Length
control is available when saving data to CSV, ASCII XY, or BIN format
files. It limits the maximum number of data points that can be saved.
When
Max Length
is enabled, the maximum number of waveform data points is
saved.
The actual number of data points saved depends on displayed data and on these
factors:
• Whether acquisitions are running. When stopped, data comes from the raw
acquisition record. When running, data comes from the smaller measurement
record.
• Whether the oscilloscope was stopped using
[Stop]
or
[Single]
. Running
acquisitions split memory to provide fast waveform update rates. Single
acquisitions use full memory.
• Whether only one channel of a pair is turned on. (Channels 1 and 2 are one pair,
channels 3 and 4 are the other.) Acquisition memory is divided among the
channels in a pair.
• Whether reference waveforms are on. Displayed reference waveforms consume
acquisition memory.
• Whether digital channels are on. Displayed digital channels consume
acquisition memory.
• Whether segmented memory is on. Acquisition memory is divided by the
number of segments.
• The horizontal time/div (sweep speed) setting. At faster settings, fewer data
points appear on the display.
• When saving to a CSV format file, the maximum number of data points is 64K.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...