Getting Started
1
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
43
17.
Analog channel
inputs
Attach oscilloscope probes or BNC cables to these BNC connectors.
With the InfiniiVision 3000T X-Series oscilloscopes, you can set the input impedance of the
analog channels to either 50
Ω
or 1 M
Ω
. See
“To specify channel input impedance"
page 82.
The InfiniiVision 3000T X-Series oscilloscopes also provide the AutoProbe interface. The
AutoProbe interface uses a series of contacts directly below the channel's BNC connector to
transfer information between the oscilloscope and the probe. When you connect a compatible
probe to the oscilloscope, the AutoProbe interface determines the type of probe and sets the
oscilloscope's parameters (units, offset, attenuation, coupling, and impedance) accordingly.
18.
Demo 2, Ground, and
Demo 1 terminals
•
Demo 2 terminal — After a
[Default Setup]
, this terminal outputs the Probe Comp signal
which helps you match a probe's input capacitance to the oscilloscope channel to which it
is connected. See
on page 36. With certain licensed
features, the oscilloscope can also output demo or training signals on this terminal.
•
Ground terminal — Use the ground terminal for oscilloscope probes connected to the Demo
1 or Demo 2 terminals.
•
Demo 1 terminal — The oscilloscope can output training signals on this and the Demo 2
terminal. See:
[Help] > Training Signals > Training Signals
.
19.
USB Host port
This port is for connecting a USB mass storage device, printer, mouse, or keyboard to the
oscilloscope.
Connect a USB compliant mass storage device (flash drive, disk drive, etc.) to save or recall
oscilloscope setup files and reference waveforms or to save data and screen images. See
, “Save/Email/Recall (Setups, Screens, Data),” starting on page 321.
To print, connect a USB compliant printer. For more information about printing see
, “Print (Screens),” starting on page 335.
You can also use the USB port to update the oscilloscope's system software when updates are
available.
You do not need to take special precautions before removing the USB mass storage device
from the oscilloscope (you do not need to "eject" it). Simply unplug the USB mass storage
device from the oscilloscope when the file operation is complete.
CAUTION:
Do not connect a host computer to the oscilloscope's USB host port.
Use the
device port. A host computer sees the oscilloscope as a device, so connect the host computer
to the oscilloscope's device port (on the rear panel). See
page 341.
There is a second USB host port on the back panel.
20.
Digital channel
inputs
Connect the digital probe cable to this connector (MSO models only). See
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...