90
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
4
FFT Spectral Analysis
•
Max Hold
— records the maximum vertical values found at each horizontal
bucket across multiple analysis cycles and uses those values to build a
waveform. This display mode is often referred to as Max Envelope.
•
Min Hold
— records the minimum vertical values found at each horizontal
bucket across multiple analysis cycles and uses those values to build a
waveform. This display mode is often referred to as Min Envelope.
•
Span
/
Center
or
Start Freq
/
Stop Freq
— this pair of softkeys let you define the
displayed frequency range. Press the softkeys to toggle between:
•
Span
/
Center
—
Span
specifies the frequency range represented by the
width of the display. Divide span by 10 to calculate the frequency scale
per division.
Center
specifies the frequency at the center vertical grid line
of the display.
•
Start Freq
/
Stop Freq
—
Start Freq
specifies the frequency at the left side of
the display.
Stop Freq
specifies the frequency at the right side of the
display.
To set desired values, tap the softkey label on screen for a keypad entry
dialog or turn the Entry knob.
•
More FFT
— displays the More FFT Settings Menu.
2
Press the
More FFT
softkey to display additional FFT settings.
•
Window
— selects a window to apply to your FFT input signal:
•
Hanning
— window for making accurate frequency measurements or for
resolving two frequencies that are close together.
•
Flat Top
— window for making accurate amplitude measurements of
frequency peaks.
•
Rectangular
— good frequency resolution and amplitude accuracy, but use
only where there will be no leakage effects. Use on self-windowing
waveforms such as pseudo-random noise, impulses, sine bursts, and
decaying sinusoids.
•
Blackman Harris
— window reduces time resolution compared to a
rectangular window, but improves the capacity to detect smaller impulses
due to lower secondary lobes.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...