122
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
5
Math Waveforms
See Also
•
Maximum/Minimum
The Maximum operator is like the Max Hold operator without the hold. The
maximum vertical values found at each horizontal bucket are used to build a
waveform.
The Minimum operator is like the Min Hold operator without the hold. The
minimum vertical values found at each horizontal bucket are used to build a
waveform.
Peak-Peak
The Peak-Peak operator is like the Maximum operator minus the Minimum
operator. At each horizontal bucket, the minimum vertical values found are
subtracted from the maximum vertical values found to build a waveform.
Max/Min Hold
The Max Hold operator records the maximum vertical values found at each
horizontal bucket across multiple analysis cycles and uses those values to build a
waveform.
The Min Hold operator is the same, except it records the minimum vertical values.
When not used in a frequency analysis domain, these functions are often referred
to as Max Envelope and Min Envelope.
Press the
Reset Count
softkey to clear the number of waveforms evaluated.
Measurement Trend
The measurement trend math function shows measurement values for a waveform
(based on measurement threshold settings) as the waveform progresses across
the screen. For every cycle, a measurement is made, and the value is displayed on
the screen for the cycle.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...