278
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
16
Mask Testing
Run Until
The Run Until softkey lets you specify a condition on which to terminate testing.
• Forever
— The oscilloscope runs continuously. However, if an error occurs the
action specified using the
On Error
softkey will occur.
• Minimum # of Tests
— Choose this option and then use the
# of Tests
softkey
to select the number of times the oscilloscope will trigger, display the
waveform(s), and compare them to the mask. The oscilloscope will stop after
the specified number of tests have been completed. The specified minimum
number of tests may be exceeded. If an error occurs the action specified
using the
On Error
softkey will occur. The actual number of tests completed
is displayed above the softkeys.
• Minimum Time
— Choose this option and then use the
Test Time
softkey to
select how long the oscilloscope will run. When the selected time has passed
the oscilloscope will stop. The specified time may be exceeded. If an error
occurs the action specified using the
On Error
softkey will occur. The actual
test time is displayed above the softkeys.
• Minimum Sigma
— Choose this option and then use the Sigma softkey to
select a minimum sigma. The mask test runs until enough waveforms are
tested to achieve a minimum test sigma. (If an error occurs the oscilloscope
will perform the action specified by the
On Error
softkey.) Note that this is a
test sigma (the max achievable process sigma, assuming no defects, for a
certain number of tested waveforms) as opposed to a process sigma (which is
tied to the amount of failures per test). The sigma value may exceed the
selected value when a small sigma value is chosen. The actual sigma is
displayed.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...