Waveform Generator
19
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
313
If the actual load impedance is different than the selected value, the displayed
amplitude and offset levels will be incorrect.
To use waveform generator logic presets
With logic level presets, you can easily set the output voltage to TTL, CMOS
(5.0V), CMOS (3.3V), CMOS (2.5V), or ECL compatible Low and High levels.
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
3
In the Waveform Generator Settings Menu, press the
Logic Presets
softkey.
4
In the Waveform Generator Logic Level Presets Menu, press one of the softkeys
to set the generated signal's Low and High voltages to logic compatible levels:
To add noise to the waveform generator output
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
3
In the Waveform Generator Settings Menu, press the
Add Noise
softkey and turn
the Entry knob to select the amount of white noise to add to the waveform
generator output.
Softkey (logic levels)
Low level
High level
TTL
0 V
+5 V (or a TTL-compatible
high level if +5 V cannot
be reached)
CMOS (5.0V)
0 V
+5 V
CMOS (3.3V)
0 V
+3.3 V
CMOS (2.5V)
0 V
+2.5 V
ECL
-1.7 V
-0.9 V
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...