Multifors 2 - Parallel Bench-top Bioreactor
Technical Data
Page
38
of
211
28 March 2018
Measuring ranges of mass flow controllers and rotamters
Max. working
volume
vessel
Measuring range MFC
mL min
-1
(Variants Standard and
High End)
Measuring range Rota-
meter
L min
-1
Air (head space)
Measuring range Rotameter
mL min
-1
CO
2
(sparger, pH control)
0.25 L
0.25 up to 25.0
0.02 up to 0.25
1.0 up to 26.0
0.50 L
0.50 up to 50.0
0.07 up to 0.64
5.0 up to 80.0
0.75 L
0.80 up to 80.0
0.06 up to 0.98
3.5.8 Antifoam
Description
Value
Sensor
Conductive with dosing needle
Control
Peristaltic pump “AF” (Antifoam)
Range
0 / 100 % (ON/OFF)
3.5.9 pH
Description
Value
Control
2 peristaltic pumps (acid, base) and/or
CO
2
valve
Range
pH 2 up to 12
Measurement accu-
racy
pH ± 0.1
Variants
Measuring system
Sensor (optional)
METTLER (analogue) Traditional pH sensor (potential measure-
ment against reference)
Manufacturer: METTLER TOLEDO
HAMILTON (digital)
Traditional pH sensor (potential measure-
ment against reference) with integrated
electronics, type Easyferm Plus ARC
Manufacturer: HAMILTON