Connection of reagent bottle to pumps:
8.
Fit both silicone hoses of the reagent bottle onto the hose con-
nector of the left pump hose (= intake side, see direction of ro-
tation of the pump in the picture on the left) on both pump
heads, e.g. both acid pumps
9.
Secure the silicone hoses with cable ties.
Connection of pumps to vessels:
10.
Fit the silicone hoses for connection with the vessels onto the
hose connector of the right pump hose (= pump side, see di-
rection of rotation of the pump in the picture on the left) of both
pump heads.