Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use
28 March 2018
Page
137
of
211
7.2.8 Preparing the Port in the Vessel Top Plate for Inoculation
The vessel top plate port must be prepared for later inoculation
with the syringe. Proceed as follows:
Procedure
1.
Loosen and remove the blanking plug from the 12 mm port us-
ing the hexagon socket spanner provided with the equipment.
2.
Ensure there is no O-ring stuck in the port, remove as neces-
sary.
3.
Insert the septum (membrane) into the port.
4.
Screw the septum collar into the port.
5.
Ensure that the blanking plug is equipped with an O-ring, fit
one as necessary.