Multifors 2 - Parallel Bench-top Bioreactor
Interferences
28 March 2018
Page
201
of
211
10.1 Interferences Drive System
Interference
Stirrer does not start.
Possible cause
Remedy
By
Parameter
Stirrer
is not activated.
Activate the parameter.
Operator
Setpoint of parameter
Stirrer
= 0.
Set setpoint > 0.
Operator
Parameter
pO
2
is activated and set
to oxygen control via the stirrer (cas-
cade).
Switch cascade off and test operation via parameter
Stirrer
.
Operator
Interference
Motor control is volatile, irregular or stops.
Possible cause
Remedy
By
PID settings of parameter Stirrer are
not correct
Reset to default settings
Operator
Interference
Unusual sounds when the stirrer is activated.
Possible cause
Remedy
By
Stirrer is in contact with other vessel
components, e.g. sensors.
Stop the bioreactor. Shutdown the system and turn
off the equipment.
Correctly mount the mounting parts of the culture
vessel while observing internal safety regulations.
Operator
Stirrer shaft is not correctly seated in
the centre bearing.