Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use
Page
146
of
211
28 March 2018
7.4 Preparations after Sterilisation
7.4.1 Mounting the Vessel Holder with the Vessels on the Basic Unit
CAUTION!
Danger of scalding and burns due to contact
with hot surfaces!
Touching the vessel holder and support frame, the
vessels or the reagent bottles straight after sterili-
sation in the autoclave carries a severe risk of
scalding and burns due to the still hot surfaces.
Therefore:
– Let the whole unit of vessel holder, vessels etc.
cool down after sterilisation.
– Always wear protective gloves.
When the unit of the vessel holder and support frame with the ves-
sels, reagent bottles etc. has cooled down sufficiently, it can be
mounted on the basic unit again:
Proceed as follows:
Procedure
1.
Ensure that the vessel retaining brackets are released i.e.
both clamps between the thermal blocks are engaged; (see
the picture on the left) release them as necessary.