Multifors 2 - Parallel Bench-top Bioreactor
Installation and Initial Operation
28 March 2018
Page
101
of
211
6.5 Connecting the Master Unit with the Satellite Unit(s)
Up to 6 bioreactors (= 3 basic units) can be operated and con-
trolled in parallel or individually using one operating panel.
This is achieved by making the following connections:
Connect the water feed and return of the master unit to the
water feed and return of each satellite unit using pressure
hoses
Connect the circuit board (controller) of the master unit with
the circuit board (controller) of each satellite unit using a iDDC
satellite cable
Depending on the configuration: connect the gassing lines of
the master unit with the gassing lines of the satellite unit(s).
NOTICE!
The gassing lines and Legris push-in connectors
are correspondingly labelled and numbered.