Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use
Page
112
of
211
28 March 2018
7.2 Preparations before Sterilisation
7.2.1 Preparing the Vessel
WARNING!
Risk of injury and loss of property due to dam-
aged vessel!
A damaged culture vessel (e.g. cracks, scratches,
etc.) can burst when under pressure. This can lead
to severe injuries and loss of property!
Therefore:
– Check the culture vessel on damages and re-
place as necessary before every use.
CAUTION!
Risk of damaging the sensor cables!
Removing the vessel from the vessel holder with-
out disconnecting the cable connections between
the different sensors and equipment or without re-
moving the temperature sensor from the pocket in
the vessel top plate can damage the sensor cables
and destruct their functionality.
Therefore:
– Disconnect all cable and hose connections be-
tween vessel and equipment, before removing
the vessel.
Disconnecting the vessel from the basic unit
To disconnect the vessel from the basic unit, proceed as follows:
Procedure
1.
Disconnect all hoses and cable connections between vessel
and basic unit.
2.
Compress both clamps of the snap closure between the ther-
mal blocks until they are engaged.
The vessel retaining brackets on the basic unit are released.