Multifors 2 - Parallel Bench-top Bioreactor
Operation
28 March 2018
Page
165
of
211
For inoculation proceed as follows:
Procedure
1.
Plug the needle onto the syringe and draw back the syringe
plunger to fill it with the necessary volume of inoculum.
2.
Unscrew and remove the blanking plug out of the septum col-
lar.
3.
Pierce the septum with the syringe needle and push down the
syringe plunger.
NOTICE!
A few drops of 70% ethanol can be placed on the
septum surface as an additional protection against
contamination.
4.
Reclose the port with the blanking plug.