Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use
28 March 2018
Page
127
of
211
Antifoam sensor with clamping adapter 12 mm ports
a) Carefully and manually screw the antifoam sensor on the
thread of the housing of the clamping screw into the port
and tighten it.
6.
Prepare the antifoam sensor for adjusting it in its height:
Antifoam sensor with clamping adapter 10 mm ports
a) Slightly loosen the screw on the adapter with a slotted
screw driver (no. 2).
Antifoam sensor with clamping adapter 12 mm ports
b) Carefully and manually loosen the hollow screw.
7.
Adjust the antifoam sensor in its required height.
Ensure that the sensor head does not touch the clamping
adapter, respectively the hollow screw on it.
NOTICE!
Mount the antifoam sensor too low rather than too
high. To pull it up during fermentation carries a
much lower risk of contamination than pushing it
down.
8.
Fixate the antifoam sensor in its height:
Antifoam sensor with clamping adapter 10 mm ports
a) Carefully tighten the screw on the clamping adapter with
the slotted screw driver.
Antifoam sensor with clamping adapter 12 mm ports
a) Carefully and manually screw the hollow screw in and
tighten it.