Multifors 2 - Parallel Bench-top Bioreactor
Installation and Initial Operation
Page
96
of
211
28 March 2018
CAUTION!
Danger of loss of property due to use of oily or
wet compressed air!
Use of oily or wet compressed air may lead to dam-
ages to the mass flow controller (option)!
Therefore:
– Only use dry, clean and oil free compressed air
The internal gas supply must meet the following conditions:
Conditions gas supply
Pressure regulation respectively reduction (2 ± 0.5 bar)
Manometer for primary pressure control
Constancy of gas supply
Only use clean, dry, oil- and dust free air/gases
Conditions hoses
Only use pressure-resistant and intact hoses.
Only use appropriate hoses, use adaptors as necessary.
Secure hoses with the appropriate clamps
6.2.3.2 Exit Gas
Ensure the exit gas is safely led away using the appropriate
pressure-resistant and intact hose.
Ensure the operating environment is provided with a ventila-
tion system, appropriate to the application.