Multifors 2 - Parallel Bench-top Bioreactor
Operation
28 March 2018
Page
167
of
211
Proceed as follows:
Procedure
1.
If present, remove the clamp from the sample hose.
2.
Check all screw connections on the valve assembly for tight-
ness. Carefully tighten the screw connections with two fingers
as necessary.
3.
If present, remove the cap.
4.
If desired, carry out a wiping disinfection.
5.
Place the Luer Lock syringe on the sample valve.
6.
Turn the syringe a quarter-turn clockwise to lock.
7.
Pull back on the syringe piston –
NOT COMPLETELY!
- in or-
der to draw up the desired sample volume.
Rinsing the dip tube with sterile air
The dip tube and its tubing line can be rinsed with sterile air after
sampling.