4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 142
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.10.1 Blood Leak Detector Limit
You can select the limit value for the alarm
activation of the blood leak detector in the
Calibration 2.7 Blood Leak Detector, BL Limit
menu.
BL Limit
Standard Limit (0.5 ml/min at haematocrite HK
0.45)
The alarm limit is set to 0.5 ml/min at a
haematocrite of 0.45, i.e. a blood leak detector
alarm is activated if the alarm limit is exceeded.
AAMI Limit (0.35 ml/min at haematocrite HK
0.25)
The alarm limit is set to 0.35 ml/min at a
haematocrite of 0.25 (AAMI standard), i.e. a blood
leak detector alarm is activated if the alarm limit
is exceeded.
Please Note
The same blood leak detector is used for both
settings:
•
The calibration procedure is identical, but with
different tolerance ranges
•
The test procedure is identical