4.8.3.11.3 Position Setting for BICP, KP and UFP
The sub-menu 2.8.2 Position Setting for BICP, KP
and UFP can be selected with the Position Setting
key.
The sub-menu 2.8.1 Stroke Value of BICP, KP and
UFP can be selected with the Stroke Value key.
The following settings are required:
BICP Parking:
180 °
KP Parking:
180 °
UFP Parking:
180 °
UFP Positive Pressure Test:
115 °
(used for DFS pressure test)
UFP Negative Pressure Test:
90 °
4.8.3.11.4 Stroke Value of BICP, KP and UFP
The sub-menu 2.8.1 Stroke Value of BICP, KP and
UFP can be selected from the menus 1.17 UF
Pump, or 2.8 Calibration Flow of BICP, KP and UFP.
The stroke values (µl/head rotation) can be
entered directly if the exact values are known for
the bicarbonate pump BICP and the concentrate
pump. The stroke value for the ultrafiltration
pump UFP should not be changed, because it is
determined by the calibration of the pump with a
balance.
It is recommended to calibrate the UFP with the
procedure described (with balance) because the
accuracy of the stroke value might be insufficient
for a proper UF control of the system.