4.8.3.10
Calibration Blood Leak Detector
You can check and calibrate the blood leak detector
BL
. The rinsing block must
have an even temperature for the calibration. If both dialyser couplings are
connected to the rinsing bridge the flow pumps
FPA/FPE
work in main flow. If
only one coupling is connected
FPE
is stopped. The reference solution can be
drawn in by the red coupling via
FPA
.
VBP
FBK2
LAFS
VLA
VDA
LA
BL
PDA
FPA
8
VDE
TSDE
T
SB2
SB1
FVD
SBS2
SBS1
LVD
VD
M
blue
red
Fig.: Rinsing Bridge with Blood Leak Detector BL
VBP
FBK2
LAFS
VLA
VDA
LA
BL
PDA
FPA
8
VDE
TSDE
T
SB2
SB1
FVD
SBS2
SBS1
LVD
VD
M
blue
red
Fig.: Rinsing Bridge with Red Coupling Immerged in Reference Solution for BL Calibration