4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 41
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Auto Switch-On by Weekly Schedule
The parameters for the switch-on and the weekly
schedule can be set in the menu Auto Switch-On
by Weekly Schedule.
The keys for Auto Switch-On with Rinsing - No
Weekly Schedule and Auto Switch-On by Weekly
Schedule are interlocked.
Start on Sunday/Start on Monday
The table for the weekly schedule can either start
with Sunday or Monday.
Weekly Schedule
The parameters for the weekly schedule can be set
in the Weekly Schedule menu.
The selected disinfection method is performed
according to the preselected Weekday and Time
parameters:
•
Monday, Tuesday ....... Sunday
•
Night (Moon Symbol), Dialog switches off
after disinfection is completed
•
Day (Sun Symbol), Dialog does not switch off
and remains in Rinsing
•
Switch-On Time
•
Disinfection Method
•
Disinfectants
The table can also be modified by the user in the
therapy/disinfection mode.
Auto Switch-On without Daily Confirmation
If YES is Activated
No confirmation required for the automatic switch-on at the end of each
disinfection.
If NO is Activated
A confirmation is required for the automatic switch-on at the end of each
disinfection.
Maximum Out of Action Time
Limit OFF
The function is activated if the Limit OFF key is not pressed.
The Limit OFF key is activated in the default setting, the parameter keys are
disabled, i.e. parameter values can not be set.