4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 42
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Disinfection Configuration Data
Chemical (long), Chemical (short)
The parameters for the disinfectants are set in the
Disinfection Configuration Data, Chemical (Long),
Chemical (Short) table. The following parameters
are available:
•
Disinfectant Name, Type and Volume
•
Reaction Time Long/Short
•
Rinsing Time
•
Operating Temperature
•
Minimum Conductivity
Details can be found in Chapter 3 repair
instructions.
Disinfection Configuration Data
Central Thermal, Central Chemical, Rinsing
The parameters for the water inlet system can be
set in the menu Disinfection Configuration Data,
Central Thermal, Central Chemical, Rinsing.
Central Thermal:
•
Inlet Flow
•
Time
Central Manual Chemical:
•
Inlet Flow
•
Inlet Time
•
Rinsing Flow
•
Rinsing Time
Central Automatic Chemical:
•
Inlet Volume
•
Switch-off for Retention without Automatic
Switch-on
•
Retention Time
•
Rinsing Flow
•
Rinsing Time
Rinsing:
•
Inlet Flow
•
Time